Dosimetric tests of an extrapolation chamber in standard computed tomography beams
DOI:
https://doi.org/10.15392/bjrs.v9i1A.1338Palabras clave:
Extrapolation chamber, computed tomography, dosimetry.Resumen
Computed tomography (CT) diagnostic exams are responsible for the highest dose values to the patients. Therefore, the radiation doses in this procedure must be accurate. For the dosimetry of CT beams, the radiation detector is usually a pencil-type ionization chamber. This type of dosimeter presents a uniform response to the incident radiation beam from all angles, which makes it suitable for such equipment since the X-ray tube executes a circular movement around the table during irradiation. However, there is no primary standard system for this kind of radiation beam yet. In order to search for a CT primary standard, an extrapolation chamber built at the Calibration Laboratory (LCI) of the Instituto de Pesquisas Energéticas e Nucleares (IPEN) was tested. An extrapolation chamber is a parallel-plate ionization chamber that allows the variation of its sensitive air volume. This chamber was used previously for low-energy radiation beams and showed results within the international recommended limits. The aim of this work is to perform some characterization tests (saturation curve, polarity effect, ion collection efficiency and linearity of response) considering the chamber depth of 1.25 mm in the radiation qualities for computed tomography beams at the LCI. The results showed to be within the international recommended limits.
Descargas
Referencias
BOONE, J.M. “The trouble with CTDI100”. Med. Phys., 34(4), 1364-1371, 2007.
SUZUKI, A.; SUZUKI, M.N. “Use of a pencil-shaped ionization chamber for measurement of exposure resulting from a computed tomography scan”. Med. Phys. 5(6), 536-9, 1978.
DIAS, S.K.; CALDAS, L.V.E. “Development of an extrapolation chamber for the calibration of beta-ray applicators”. IEEE Trans. Nucl. Sci. 45, 1666-1669, 1998.
DIAS, S.K.; CALDAS, L.V.E. “Extrapolation chamber response in low-energy radiation stand-ard beams”. J. Appl. Phys. 89, 669-671, 2001.
NEVES, L.P.; SILVA, E.A.B.; PERINI, A.P.; MAIDANA, N.L.; CALDAS, L.V.E. “Characterization of an extrapolation chamber for low-energy X-rays: Experimental and Monte Carlo preliminary results”. Appl. Radiat. Isot. 70, 1388-1391, 2012.
IEC - INTERNATIONAL ELECTROTECHNICAL COMMISSION: Medical diagnostic X-ray equipment. Radiation conditions for use in the determination of characteristics. Report 61267, 2nded. IEC Genève, 2005.
IEC - INTERNATIONAL ELECTROTECHNICAL COMMISSION: Medical electrical equipment-dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging. Report 61674, IEC Genève, 1997.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/