Pile-up correction for coincidence counting using a CAEN 1724 digitizer

Autores

  • Guilherme Soares Zahn Instituto de Pesquisas Energéticas e Nucleares (IPEN/CNEN-SP)
    • Iberê Souza Ribeiro Jr Instituto de Pesquisas Energéticas e Nucleares (IPEN/CNEN-SP)
      • Frederico Antonio Genezini Instituto de Pesquisas Energéticas e Nucleares (IPEN/CNEN-SP)

        DOI:

        https://doi.org/10.15392/bjrs.v7i2A.578

        Palavras-chave:

        CAEN 1724, pile-up correction, coincidence

        Resumo

        In conventional gamma-ray spectrometry, the probability of pile-up effects is considered to be proportional to the dead-time, and is usually neglected for low dead-times (below 4-5%). In gamma-gamma coincidence spectrometry, though, while the dead time takes into account only events that are actually digitized, the pile-up effects are proportional to the actual gamma-ray detection rate in each detector, not only to the ones that trigger the coincidence gate. Thus, the pile-up corrections may not be so easy to assess as in single spectrometry systems. In this work, a system composed of two HPGe detectors coupled to a CAEN v1724 digitizer is studied. A 3kBq 60Co source was analyzed, both alone and in the presence of other radioactive sources (137Cs, 133Ba and 152Eu), and the resulting coincidence peak areas were compared to assess the effectiveness of two distinct corrections: a simple normalization by the live time of acquisition and the normalization by the count rate obtained using a pulse generator. The results obtained stress the need to use the pulse generator in this specific setup in order to get accurate results.

        Downloads

        Os dados de download ainda não estão disponíveis.

        Referências

        CAEN MC2Analyzer User Manual: Software for digital Multi Channel Analyzer. Viareggio, Italy, 2017. Available at: http://www.caen.it/servlet/checkCaenManualFile?Id=12608. Last Ac-cessed: 07 Nov. 2017.

        KNOLL, G. F. Radiation Detection and Measurement, Hoboken: John Wiley & Sons, 2011.

        MASCARENHAS, P. H. A.; GENEZINI, F. A.; ZAHN, G. S., Efficiency stability in HPGe detec-tors, In: XXXVIII REUNIÃO DE TRABALHO SOBRE FÍSICA NUCLEAR NO BRASIL, 2015, Mangaratiba, Proceedings. São Paulo, Sociedade Brasileira de Física, 2015.

        WAPSTRA, A. The coincidence method. In: SIEGBAHN, K. Alpha-, Beta- and Gamma-Ray Spectroscopy, Amsterdam: North-Holland, 1965. p. 539-555.

        WIERNIK, M. Comparison of several methods proposed for correction of dead-time losses in the gamma-ray spectrometry of very short-lived nuclides. Nucl. Instrum. Meth. v. 95, p.13-18, 1971.

        YOHO, M.; LANDSBERGER, S. Determination of Selenium in coal fly ash via gamma-gamma coincidence neutron activation analysis. J. Radioanal. Nucl. Chem. v. 307, p.733-737, 2016

        Downloads

        Publicado

        07-02-2019

        Edição

        Seção

        The Meeting on Nuclear Applications (ENAN)