Influence of the volume of laboratory-developed simulators on the counting efficiency (kCPS/MBq) of the PET scanner belonging to LIM/CDTN

Abstract: This study investigated the influence of phantom volume on the conversion coefficient between count rate (CPS) and activity (MBq) for a preclinical Positron Emission Tomography (PET) scanner, the Triumph® LabPET Solo 4, installed at the LIM/CDTN (Laboratório de Imagem Molecular do Centro de Desenvolvimento da Tecnologia Nuclear). Three cylindrical phantoms (S, M, and L) with different internal volumes (17.7 mL, 22.0 mL, and 26.5 mL, respectively) were fabricated. PET images were acquired using two approaches: spatial and temporal. The spatial acquisition mode was used to verify the quality and homogeneity of the reconstructed PET images for the different phantom volumes. In this analysis, no relevant variations in image quality were observed among the phantoms. The temporal acquisition mode was applied to evaluate the behavior of count rates as a function of activity, analyzing total, true, random, scattered, and noise-equivalent count (NEC). The results demonstrated a progressive reduction in counting efficiency with increasing phantom volume, indicating an inverse relationship between phantom size and detection efficiency. These findings emphasize the importance of considering phantom geometry in calibration and quantitative procedures in preclinical PET imaging studies to ensure accurate measurements. Read full article.