Comparative evaluation of the performance of thin diodes used as on-line dosimeters in radiation processing applications
DOI:
https://doi.org/10.15392/bjrs.v9i1A.1327Resumo
In this work, we report a comparison between the performance of two samples of commercial PIN photodiodes (SFH206K from Osram and S2506-04 from Hamamatsu) mainly addressing the variation of their current sensitivities with accumulated dose ranging from 0-15 kGy. All the results so far obtained have revealed that the radiation induced currents are linearly dependent on dose rates from 3.65 to 55.64 Gy/h. The current sensitivity of both unirradiated diodes (0.178 nA.h/Gy.mm3) slightly decreases with accumulated dose, namely 0.32%/kGy (SFH206K) and 1.4%/kGy (S2506-04). Although the SFH206K device compares favorably with the S2506-04, both diodes can be considered as a low budge alternative, good enough for on-line dosimetry applications in the field of radiation processing.
Downloads
Referências
DIXON, R. L.; EKSTRAND, K. E. Silicon diode dosimetry. Int J Appl Radiat Isot, v. 33, p. 1171-1176, 1982.
RIKNER, G.; GRUSSEL, E. General specifications for silicon semiconductors for use in radiation dosimetry. Phys Biol, v. 32(9), p. 1109-1117, 1987.
SAINI, A. S.; ZHU, T. C. Dose rate and SDD dependence of commercially available diode detectors. Med Phys, v. 31(4), p. 914-924, 2004. http://doi.org/10.1118/1.1650563.
CASATI, M.; BRUZZI, M.; BUCCIOLINI, M.; MENICHELLI, D.; SCARINGELLA, M.; PIEMONTE, C.; FRETWURST, E. Characterization of standard and oxygenated float zone Si diodes under radiotherapy beams. Nucl Instrum Methods A, v. 552, p.158-162, 2005. http://doi.org/10.1016/j.nima.2005.06.025.
GRIESSBACH, I.; LAPP, M.; BOHSUNG, J.; GADEMANN, G.; HARDER, D. Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams. Med Phys, v. 32(12), p. 3750-3754, 2005. http://doi.org/10.1118/1.2124547.
BRUZZI, M.; BUCCIOLINI, M.; CASATI, M.; MENICHELLI, D.; TALAMONTI, C.; PIEMONTE, C.; SVENSSON, B. G. Epitaxial silicon devices for dosimetry applications. Appl Phys Lett, v. 90, p. 172109, 2007. http://doi.org/10.1063/1.2723075.
GONÇALVES, J. A. C.; PEREIRA, L. N.; POTIENS, M. P. A.; VIVOLO, V.; BUENO, C. C. Evaluation of epitaxial silicon diodes as dosimeters in X-ray mammography. Radiat Meas, v. 71, p. 384-388, 2014. http://doi.org/10.1016/j.radmeas.2014.07.014.
SANTOS, T. C.; NEVES-JUNIOR, W. F. P.; GONÇALVES, J. A. C.; HADDAD, C. M. K.; HARKONEN, J.; BUENO, C. C. Characterization of miniature rad hard silicon diodes as dosimeters for small fields of photon beams used in radiotherapy. Radiat Meas, v. 71, p. 396-401, 2014. http://doi.org/10.1016/j.radmeas.2014.08.002.
NASCIMENTO, C. R.; ASFORA, V. K.; GONÇALVES, J. A. C.; KHOURY, H. J.; BARROS, V. S. M.; KALIL, L. F.; BUENO, C. C. The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry. Appl Radiat Isot, v. 141, p. 112-117, 2018. http://doi.org/10.1016/j.apradiso.2018.07.002.
GONÇALVES, J. A. C.; MANGIAROTTI, A.; BUENO, C. C. Current response stability of a commercial PIN photodiode for low dose radiation processing applications. Rad Phys Chem, v. 167, p. 108276-9, 2020. https://doi.org/10.1016/j.radphyschem.2019.04.026.
ISO - International Standardization Organization. Standard Practice for Dosimetry in a Gamma Facility for Radiation Processing. ISO/ASTM 51702:2013(E), 3rd ed., 2013.
ISO - International Standardization Organization. Standard Practice for Dosimetry in Radiation Processing. ISO/ASTM 52628:2013(E), 1st ed., 2013.
OSVAY, M.; TÁRCZY, K. Measurement of -dose rates by n- and p-type semiconductor detectors. Phys Stat Sol, v. (A)27, p. 285-290, 1975.
Downloads
Publicado
Edição
Seção
Licença
Direitos autorais (c) 2021 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/