Low-Dose Gamma Radiation Dosimetry Using a 60Co Irradiator: Experimental Validation and Reproducibility

Authors

  • Walewska Gouveia Silva Centro de Desenvolvimento da Tecnologia Nuclear image/svg+xml
    • Conceptualization
    • Formal Analysis
    • Investigation
    • Methodology
    • Writing – Original Draft Preparation
    • Writing – Review & Editing
  • Amir Zacarias Mesquita Centro de Desenvolvimento da Tecnologia Nuclear image/svg+xml
    • Conceptualization
    • Formal Analysis
    • Project Administration
    • Resources
    • Supervision
    • Validation
    • Writing – Review & Editing
  • Daniel Silva Calheiro Centro de Desenvolvimento da Tecnologia Nuclear image/svg+xml
    • Conceptualization
    • Funding Acquisition
    • Methodology
    • Project Administration
    • Software
    • Resources
    • Writing – Review & Editing
  • Leonardo Yago Zwetsch Varella Centro de Desenvolvimento da Tecnologia Nuclear image/svg+xml
    • Conceptualization
    • Validation
    • Visualization
  • Mateus de Jesus Silva Fluminense Federal University image/svg+xml
    • Conceptualization
    • Project Administration
    • Visualization
    • Validation
    • Supervision
  • Antônio Carlos Pinheiro de Oliveira Federal University of Minas Gerais image/svg+xml
    • Conceptualization
    • Formal Analysis
    • Project Administration
    • Supervision

DOI:

https://doi.org/10.15392/2319-0612.2026.3055

Keywords:

Dosimetry, gamma radiation, low-dose irradiation, cobalt-60

Abstract

Abstract: Low-dose ionizing radiation has been increasingly investigated in preclinical research due to its potential to modulate biological responses while minimizing the risk of significant tissue damage. However, the scientific reliability of studies conducted in the milliGray dose range strongly depends on the availability of rigorous, reproducible, and traceable dosimetric protocols. Small variations in irradiation geometry, exposure time, or shielding conditions may result in substantial uncertainties when low doses are involved, reinforcing the importance of precise dosimetric characterization and standardized experimental methodologies. In this context, the present study describes the development and validation of a low-dose gamma radiation dosimetry protocol (~41.3 mGy) implemented using a multipurpose cobalt-60 (60Co) irradiator at the Gamma Irradiation Laboratory (LIG) of the Nuclear Technology Development Center (CDTN), Brazil. The protocol was designed to support preclinical experiments, with emphasis on accuracy, reproducibility, and strict compliance with radiological protection requirements. Dosimetric planning was based on the Beer–Lambert exponential attenuation law, applied to the dimensioning of lead shielding and to the definition of the transmitted beam fraction required to achieve the target dose. These theoretical calculations were integrated with the geometric configuration of the irradiation system, including source-to-sample distance and exposure time, to ensure controlled dose delivery within the milliGray range. A conservative experimental configuration was adopted, employing the maximum operational distance available in the irradiator to reduce dose rates and improve temporal control during short irradiation times. Lead shielding elements were used to stabilize the irradiation setup and to ensure consistent attenuation of the gamma beam. All irradiation procedures followed the routine operational practices of the LIG/CDTN, ensuring dosimetric traceability and adherence to national radiological protection standards. The results demonstrate that the proposed protocol enables the controlled and reproducible delivery of low gamma doses with suitable radiation field uniformity for preclinical applications. Overall, the methodology provides a robust and adaptable dosimetric framework for experimental studies requiring low-dose gamma irradiation, contributing to improved standardization, safety, and comparability in preclinical dosimetry research.

Downloads

Download data is not yet available.

References

[1] MLADENOVA, V. et al. Damage Clustering after Ionizing Radiation and Consequences in the Processing of Chromatin Breaks. Molecules, [s. l.], v. 27, n. 5, p. 1540, 2022. DOI: https://doi.org/10.3390/molecules27051540

[2] KAUL, D. et al. Radiation Therapy in Alzheimer’s Disease: A Systematic Review. International Journal of Radiation Oncology - Biology - Physics, [s. l.], v. 119, n. 1, p. 23-41, 2024. DOI: https://doi.org/10.1016/j.ijrobp.2023.11.044

[3] MOTHERSILL, C.; SEYMOUR, C. Radiation Hormesis and Dose Response: Are Our Current Concepts Meaningful or Useful? Current Opinion in Toxicology, [s. l.], v. 29, p. 99-106, 2022. DOI: https://doi.org/10.1016/j.cotox.2022.02.008

[4] MANJUNATHA et al. Experimental Investigation of Gamma-Ray Interaction Parameters and Buildup Factors in Lanthanide Compounds: Insights into Penetration Depth. Applied Radiation and Isotopes, [s. l.], v. 212, p. 111466, out. 2024. DOI: https://doi.org/10.1016/j.apradiso.2024.111466

[5] ÖZDOGAN, H. et al. Investigation of Gamma Ray Shielding Characteristics of Binary Composites Containing Polyester Resin and Lead Oxide. Polymers, Basel, v. 16, n. 23, p. 3324, nov. 2024. DOI: https://doi.org/10.3390/polym16233324

[6] BÜYÜKYILDIZ, M. et al. Gamma-ray Attenuation Properties of Some Heavy Metal Ferroalloys for Potential Radiation Shielding Applications. Annals of Nuclear Energy, [s. l.], v. 192, p. 110060, 2024. DOI: https://doi.org/10.1016/j.pnucene.2024.105382

[7] CHAIPHAKSA, W. et al. Experimental, Theoretical and Monte Carlo Simulation Study of Radiation Shielding Properties of La₂O₃-Added Lithium Borate Glasses. Radiation Physics and Chemistry, [s. l.], v. 215, p. 111263, 2024. DOI: https://doi.org/10.1016/j.radphyschem.2024.111994

[8] INTERNATIONAL ATOMIC ENERGY AGENCY. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water (TRS-398, Rev. 1). Vienna: IAEA, 2024. DOI: https://doi.org/10.61092/iaea.ve7q-y94k

[9] INTERNATIONAL ATOMIC ENERGY AGENCY. Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards. Vienna: IAEA, 2014. (IAEA Safety Standards Series No. GSR Part 3).

[10] ISO/ASTM INTERNATIONAL. ISO/ASTM 51261:2013: Practice for Calibration of Routine Dosimetry Systems for Radiation Processing. Geneva: ISO, 2013.

[11] JOINT COMMITTEE FOR GUIDES IN METROLOGY. Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement (GUM). Sèvres: BIPM, 2008. (JCGM 100:2008).

[12] PEROCHEAU, D. et al. Optimized and Simplified Technique for the Production and Culture of Precision-Cut Liver Slices. J. Vis. Exp., [s. l.], n. 213, e67202, 2024. DOI: https://doi.org/10.3791/67202

Downloads

Published

2026-10-09

Issue

Section

Original Articles