Analysis of different protocols using an Exradin A12 cylindrical chamber for electron beam reference dosimetry

Authors

  • Andre Gonçalves Prospero Hospital das Clínicas da Faculdade de Medicina de Botucatu (HCFMB) - Setor de Radioterapia https://orcid.org/0000-0001-5163-8280
  • Ana Carolina Ferreira University Hospital of the Botucatu Medical School (HCFMB), UNESP, Department of Radiotherapy, 18618-687, Botucatu, São Paulo, Brazil
  • Diana Rodrigues Pina bUniversity Hospital of the Botucatu Medical School (HCFMB), UNESP, Department of Tropical Diseases and Imaging Diagnosis, 18618-687, Botucatu, São Paulo, Brazil
  • Antonio Cesar Pernomian University Hospital of the Botucatu Medical School (HCFMB), UNESP, Department of Radiotherapy, 18618-687, Botucatu, São Paulo, Brazil
  • Wander de Oliveira University Hospital of the Botucatu Medical School (HCFMB), UNESP, Department of Radiotherapy, 18618-687, Botucatu, São Paulo, Brazil

DOI:

https://doi.org/10.15392/2319-0612.2023.2232

Keywords:

electron beam, Dosimetry, cylindrical chambers, parallel plate chambers

Abstract

The Technical Report Series 398 (TRS-398), Electron Dosimetry Working Party (EWDP), and Task Group 51 (TG 51) are the most important protocols for reference dosimetry. In the case of electron beam reference dosimetry, these protocols recommend using parallel-plate ionization chambers for beams with R50 values below specific thresholds. However, recent papers suggested using cylindrical chambers for reference dosimetry of all electron beam energies. Here we compared different protocols using a cylindrical chamber with the recommendations of using a parallel-plate chamber and the TRS-398 formalism for the dosimetry of several electron beam energies.  We employed electron beams with nominal energies of 4, 6, 9, 12, and 15 MeV of a Varian 2100C linear accelerator, an Exradin A12, and an Exradin P11 chamber for the analysis. The results showed differences below 3% when comparing the cylindrical chamber and alternative protocols with the parallel-plate chamber and the TRS-398 formalism for electron beams reference dosimetry. These results can bring confidence in using a cylindrical chamber for reference electron beam dosimetry, which can make the electron beam dosimetry procedure simpler and faster.

Downloads

Download data is not yet available.

References

IAEA (INTERNATIONAL ATOMIC ENERGY AGENCY) – Technical Report Series 398: Absorbed Dose Determination in External Beam Radiotherapy, Vienna: INTERNATIONAL ATOMIC ENERGY AGENCY, 2001. Available at: <https://www.iaea.org/publications/5954/absorbed-dose-determination-in-external-beam-radiotherapy>. Last accessed: 30 Apr. 2023.

ALMOND, P. R.; BIGGS, P J.; COURSEY, B. M.; HANSON, W. F.; HUQ, M. S.; NATH, R.; ROGERS, D. W. O. AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beams. Med Phys, v. 26(9), p. 1847-1870, 1999. Available at <https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.598691> . Last accessed: 30 Apr. 2023.

THWAITES, D. I.; DUSAUTOY, A.R.; JORDAN, T.; MCEWEN, M. R.; NISBET, A.; NAHUM, A. E.; PITCHFORD, W. G. The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration. Phys Med & Biol, v. 48(18), p. 2929-2970, 2003. Available at: <https://iopscience.iop.org/article/10.1088/0031-9155/48/18/301>. Last accessed: 30 Apr. 2023.

WITTKAMPER, F. W.; THIERENS, H.; VAN DER PLAETSEN, A.; WAGTER, C.; MIJNHEER, B. J. Perturbation correction factors for some ionization chambers commonly applied in electron beams. Phys Med & Biol, v. 36(12), p. 1639-1652, 1991. Available at: <https://iopscience.iop.org/article/10.1088/0031-9155/36/12/008>. Last accessed: 30 Apr. 2023.

MUIR, B.R. and M.R. McEWEN. On the use of cylindrical ionization chambers for electron beam reference dosimetry. Med Phys, v. 44(12), p. 6641-6646, 2017. Available at: <https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.12582>. Last accessed: 30 Apr. 2023.

STEWART, K. and SEUNTJENS, J. Comparing calibration methods of electron beams using plane‐parallel chambers with absorbed‐dose to water based protocols. Med Phys, v. 29(3), p. 284-289, 2002. Available at: <https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.1449876>. Last accessed: 30 Apr. 2023.

ZINK, K.; CZARNECKI, S.; LOOE, H. K.; VON VOIGTS-RHETZ, P.; HARDER, D. Monte Carlo study of the depth‐dependent fluence perturbation in parallel‐plate ionization chambers in electron beams. Med Phys, v. 41(11), p. 111707, 2014. Available at: <https://aapm.onlinelibrary.wiley.com/doi/full/10.1118/1.4897389>. Last accessed: 30 Apr. 2023.

BUCKLEY, L.A. and ROGERS, D. W. O. Wall correction factors, for parallel‐plate ionization chambers. Med Phys, v. 33(6Part1), p. 1788-1796, 2006. Available at: <https://aapm.onlinelibrary.wiley.com/doi/10.1118/1.2161403>. Last accessed: 30 Apr. 2023.

MUIR, B. R. and ROGERS, D. W. O. Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types. Med Phys, v. 41(11), p. 111701, 2014. Available at: <https://aapm.onlinelibrary.wiley.com/doi/full/10.1118/1.4893915>. Last accessed: 30 Apr. 2023.

MUIR, B. R., MCEWEN, M. and ROGERS, D. W. O. Beam quality conversion factors for parallel‐plate ionization chambers in MV photon beams. Med Phys, v. 39(3), p. 1618-1631, 2012. Available at: <https://aapm.onlinelibrary.wiley.com/doi/full/10.1118/1.3687864>. Last accessed: 30 Apr. 2023.

MUIR, B. R. A modified formalism for electron beam reference dosimetry to improve the accuracy of linac output calibration. Med Phys, v. 47(5), p. 2267-2276, 2020. Available at: <https://aapm.onlinelibrary.wiley.com/doi/full/10.1002/mp.14048>. Last accessed: 30 Apr. 2023.

INCA (INSTITUTO NACIONAL DO CANCER) TEC DOC - 1151: aspectos físicos da garantia da qualidade em radioterapia. Programa de Qualidade em Radioterapia, Rio de Janeiro, Brasil Ministério da Saude, 2000. Available at: <https://www.inca.gov.br/publicacoes/notas-tecnicas/tecdoc-1151-aspectos-fisicos-da-garantia-da-qualidade-em-radioterapia>. Last accessed: 30 Apr. 2023.

Downloads

Published

2023-06-21

Issue

Section

Articles

How to Cite

Analysis of different protocols using an Exradin A12 cylindrical chamber for electron beam reference dosimetry. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 11, n. 2, p. 1–16, 2023. DOI: 10.15392/2319-0612.2023.2232. Disponível em: https://www.bjrs.org.br/revista/index.php/REVISTA/article/view/2232.. Acesso em: 6 may. 2024.

Similar Articles

1-10 of 302

You may also start an advanced similarity search for this article.