Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner

Authors

DOI:

https://doi.org/10.15392/bjrs.v10i3.2006

Keywords:

quality control, NEMA NU 4-2008, Preclinical PET

Abstract

This study aimed to evaluate the general performance of the preclinical PET scanner (LabPET Solo 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 protocol. The performance parameters include: (i) spatial resolution; (ii) scatter fraction and counts rate performance; (iii) sensitivity and (iv) image quality. For the evaluation specific simulators and a sealed radioactive source of Sodium-22 were used. The results revealed that the NEMA-based radial spatial resolution ranged from 1.79 mm at the center of the FOV to 2.49 mm at a radial offset of 25mm. Tangential resolution ranged from 1.95mm at the center of the FOV to 3.50mm at a radial offset of 25mm. Axial resolution ranged from 1.78mm at the center of the FOV to 5.42mm at a radial offset of 25mm. Iterative image reconstruction improved the radial spatial resolution to 0.93mm at the center of the FOV. For the mouse-sized phantom, the peak noise equivalent count rate (NECR) was 37.8kcps at 85MBq whereas the peak true count rate was 169.9kcps at 125MBq with a count loss of 7.8% for the maximum rate reached at 103.3MBq. The peak of sensitivity (S) and absolute sensitivity (SA) at the center of axial field-of-view (FOV) is 5.7cps.Bq-1 and 0.63%, respectively. The overall imaging capabilities of the scanner were assessed using the NEMA image-quality phantom. The image roughness was 9.5%. Spillover rates (SORs) in air and in water were 0.26 and 0.17 respectively. The recovery coefficients (RCs) were 0.12 for the 1-mm-diameter rod and 0.91 for the 5-mm-diameter rod. The results demonstrate that CDTN/CNENsmall animal PET scanner has adequate reliability and acquired images are compatible with international quality standards for the molecular imaging research.

Downloads

Download data is not yet available.

References

YAO R.; LECOMTE R.; CRAWFORD E. Small-Animal PET: What is it, and why do we need it? Journal of Nuclear Medicine Technology. Vol. 40 nº3, pp.157-165. 2012. DOI: https://doi.org/10.2967/jnmt.111.098632

GONTIJO, R. M. G. Proposta de Programa de Garantia da Qualidade para Imagem Mole-cular Pré-Clínica. 2019. 275 f. Doctoral Thesis (Doutorado em Ciência e Tecnologia das Radi-ações, Minerais e Materiais) – Centro de Desenvolvimento da Tecnologia Nuclear, Comissão Nacional de Energia Nuclear, Belo Horizonte, In Portuguese, 2019.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Small Animal Positron Emission Tomographs. Rosslyn VA; 2008 Standards Publication NU 4-2008.

GONTIJO, R. M. G., FERREIRA, A. V., SILVA, J. B., MAMEDE, M. Quality control of small animal PET scanner: The Brazilian Scenario. Brazilian Journal of Radiation Sciences. vol. 08 (2). pp. 1-09. 2020.

GE Healthcare Technologies, 2011. Triumph Service Guide Technical Publication. Revision Draft 6, Copyright. Available in: <http://www.gehealthcare.com> 2011.

TETRAULT, M.A. et al., 2008. System Architecture of the LabPET Small Animal PET Scanner. IEEE Transactions on Nuclear Science. 55(5), 2546-2550. DOI: https://doi.org/10.1109/TNS.2008.2002326

FONTAINE, R. et al. 2009. The hardware and signal processing architecture of LabPETTM, a small animal APD-based digital PET scanner. IEEE Transactions on Nuclear Science. 56, 3-9. DOI: https://doi.org/10.1109/TNS.2008.2007485

GONTIJO, R. M. G., FERREIRA, A. V., SILVA, J. B., MAMEDE, M. Image quality assess-ment using NEMA NU 4/2008 standards in small animal PET scanner. Brazilian Journal of Radiation Sciences. 07 (2A). pp. 1-13. 2019.

Bergeron, M. et al. Imaging performance of LabPET APD-based digital PET scanners for pre-clinical research. Physics in medicine and Biology. 59, 661-678. 2014. DOI: https://doi.org/10.1088/0031-9155/59/3/661

Prasad, R., Ratib, O., Zaid, H. NEMA NU-04-based Performance Characteristics of the LabPET-8TM Small Animal PET Scanner. Physics in Medicine and Biology. 56(20), 6649-6664. 2011. DOI: https://doi.org/10.1088/0031-9155/56/20/009

Bergeron, M. et al. Performance Evaluation of the LabPET APD-Based Digital PET Scan-ner. IEEE Transactions on Nuclear Science. 56(1). 2009. DOI: https://doi.org/10.1109/TNS.2008.2010257

Gontijo, R.M.G., Ferreira, A.V., Souza, G.C.A., Mendes, B.M., Silva, J.B., Mamede, M. 2020. Image quality evaluation of a small animal PET scanner. Brazilian Journal of Radiation Sciences. 8(1), 01-13. 2020. DOI: https://doi.org/10.15392/bjrs.v8i1.1155

IAEA - International Atomic Energy Agency (IAEA). Quality Control Guidance for Nu-clear Medicine Equipment. Guidelines of Radiation and Nuclear Safety Authority – STUK, 2010.

GONTIJO, R. M. G., FERREIRA, A. V., SOUZA, G. C. A., BARBOSA, J. V. C., MAMEDE, M. Current Brazilian Scenario about Quality Assurance in preclinical PET imaging systems. 2021 International Nuclear Atlantic Conference – INAC 2021 Virtual meeting, Brazil, November 29 – December 2, 2021.

OSBORNE, D. R., KUNTNER, C., BERR, S., STOUT, D. Guidance for Efficient Small Animal Imaging Quality Control. Molecular Imaging and Biology. 19 pp. 485-498. 2016. DOI: https://doi.org/10.1007/s11307-016-1012-3

Gontijo, R.M.G., Ferreira, A. V., Silva, J. B., Mamede, M. Image quality assessment using NEMA NU 4/2008 standards in small animal PET scanner. Brazilian Journal of Radiation Sciences. 7(2A), 01-12. 2019. DOI: https://doi.org/10.15392/bjrs.v7i2A.580

Downloads

Published

2022-09-18

Issue

Section

Articles

How to Cite

Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 10, n. 3, 2022. DOI: 10.15392/bjrs.v10i3.2006. Disponível em: https://www.bjrs.org.br/revista/index.php/REVISTA/article/view/2006.. Acesso em: 5 may. 2024.

Similar Articles

1-10 of 548

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)

1 2 > >>